Section 5 Command Descriptions
5.19
COUNT
The
COUNT
command returns the loop count, appended with the serial output
terminator(s), of the last invoked Macro buffer. The number returned is the remaining
number of loops left to execute. It, therefore, is a decrementing count. If Macro 1
requests a
COUNT
then calls Macro 0, any subsequent
COUNT
requests made by Macro
1 will return the Macro 0 loop
COUNT
. The following contents of Macro 1's buffer
should illustrate this
MACRO 1
Creates Macro Buffer #1
COUNT
This will return Macro #1's loop count
DOMACRO 0
This command executes Macro #0
COUNT
This will return Macro #0's loop count since it
was the last Macro to be invoked
ENDM01
If Macro #1, in the previous example, was invoked multiple times, each time the
loop counts will be reported as described. This is due to the fact that looping is a re-
invocation of the looped Macro. This command is only valid when contained within a
Macro. Execution outside of a Macro will generate an '
INVALID
COMMAND
' error.
SYNTAX COUNT
RESPONSE numeric loop count (
1
to
255
) of last invoked Macro buffer.
MODE Any
BUS STATES None
EXAMPLE
10 PRINT #1,"MACRO"
Build Macro #0
20 PRINT #1,"COMMENT 'Loop Number =\'"
30 PRINT #1,"COUNT"
40 PRINT #1,"ENDM"
50 PRINT #1,"DOMACRO0,5" Execute the Macro five times
60 FOR N = 1 TO 5
70 INPUT #1,L$:PRINT L$
Read Comment & Count
80 NEXT N