© 1998 by IOtech, Inc.
December 1999
Printed in the United States of America
WavePort/312P
*
User’s Manual
Power Measurement, Display, Storage, & Analysis Instrument
with EasyPower
Measure
Software
IOtech, Inc.
25971 Cannon Road
Cleveland, OH 44146-1833
Phone: (440) 439-4091
Fax: (440) 439-4093
E-mail: sales@iotech.com
Internet: http://www.iotech.com
*WavePort/312P was formerly known as PowerVista/312. PowerVista® is a registered trademark of PowerVista
Software. With exception of this cover page, the WavePort/312P product name has not yet been incorporated into this
document.
WavePort/312P
User’s Manual
p/n
1006-0901
Rev 1.0
Original release 8/98 per EO# 2081R3
Zobrazit stránku 0
1 2 3 4 5 6 ... 81 82

Shrnutí obsahu

Strany 1 - WavePort/312P

© 1998 by IOtech, Inc. December 1999Printed in the United States of AmericaWavePort/312P* User’s ManualP

Strany 2

1-4 Introduction and Hardware Setup PowerVista/312 User’s Manual1.1.3 Frequently Asked QuestionsCan I Operate PowerVista/312 with the Unit Close

Strany 3 - PowerVista/312 User’s Manual

PowerVista/312 User’s Manual Introduction and Hardware Setup 1-51.1.4 Mounting the LaptopFigure 1-3. Attaching a Laptop to the Cover PlateIt is

Strany 4 - Table of Contents

1-6 Introduction and Hardware Setup PowerVista/312 User’s Manual1.1.5 Connecting the PCThis section involves connecting your laptop to the acqui

Strany 5

PowerVista/312 User’s Manual Install Software and Change PC Settings 2-12 Install Software and Change PC Settings2.1 Introduction to EasyPower M

Strany 6

2-2 Install Software and Change PC Settings PowerVista/312 User’s ManualFigure 2-1. The Control Panel Power Management ICON.Once activated, go th

Strany 7 - 1.1 What is PowerVista/312 ?

PowerVista/312 User’s Manual Install Software and Change PC Settings 2-3Figure 2-6. Default Screen Saver and Energy Management in Win95.2.6 Disable

Strany 8

2-4 Install Software and Change PC Settings PowerVista/312 User’s ManualFigure 2-4. Virtual Memory Management.2.7 Disable Additional Process Tim

Strany 9

PowerVista/312 User’s Manual Software Framework 3-13 Software Framework3.1 OverviewEasyPower Measure is a Windows multiple document interface

Strany 10 - What Does the UPS Option Do?

3-2 Software Framework PowerVista/312 User’s Manual3.2 File AccessFile access for EasyPower Measure follows standard Windows convention for a mult

Strany 11 - 1.1.4 Mounting the Laptop

PowerVista/312 User’s Manual Software Framework 3-33.2.4 File CloseFile Close terminates the presently selected Acquisition Widow. This is simil

Strany 13 - 2.3 Disable Power Management

3-4 Software Framework PowerVista/312 User’s ManualTable 3. Supplied User Configuration Files.Configuration File Comment120VdemY.cfg 120 V, Wye c

Strany 14

PowerVista/312 User’s Manual Software Framework 3-53.4.1 Print Print either a result summary or a single result window to the output device sele

Strany 15

3-6 Software Framework PowerVista/312 User’s Manual3.5.2 Paste Shots The Paste Shots command accesses the list of shots to copy stored in memory

Strany 16

PowerVista/312 User’s Manual Software Framework 3-73.7 ToolsSeveral tools have been supplied to make accessing results simple and convenient. Wit

Strany 17 - 3 Software Framework

3-8 Software Framework PowerVista/312 User’s Manual3.7.3 Cycle Graph Backward Cycle Graph Backward causes the graphical and tabulated text resul

Strany 18 - 3.2 File Access

PowerVista/312 User’s Manual Software Framework 3-9Table 5. Shot Four Letter Pneumonics for Database View.Shot Type Shot Type PneumonicPhasor Dia

Strany 19 - 3.3 User Configuration

3-10 Software Framework PowerVista/312 User’s Manual3.8.2 CascadeCascade will auto-arrange all Acquisition Windows in a folder arrangement where

Strany 20 - 3.4 Printing

PowerVista/312 User’s Manual Software Framework 3-11Figure 3-5. Acquisition Windows Tiled Vertically.3.8.5 Arrange IconsIf any of the Applicatio

Strany 21 - 3.5 Copy - Paste Shots

3-12 Software Framework PowerVista/312 User’s Manual3.9.1 Main ToolbarAs of Version 1.0, the Main Toolbar incorporates the following direct acces

Strany 22 - 3.6 Copy to Clipboard

PowerVista/312 User’s Manual Software Framework 3-13Each of the toolbar buttons can also be set in the Configuration / System dialog.See the Phasi

Strany 23 - 3.7 Tools

PowerVista/312 User’s ManualiWarrantyPowerVista/312’s warranty is as stated on the enclosed warranty card. No other warranty is expressed orimplied.

Strany 24 - 3.7.6 Take Snapshot

3-14 Software Framework PowerVista/312 User’s Manual⇒ Event/Demand Capture FIFO Overruns: should not be allowed. As noted above in Cycle-by-Cyc

Strany 25 - 3.8 Window Management

PowerVista/312 User’s Manual Software Framework 3-15• Left Click and Drag of Mouse Over Shot List: selects and highlights a list of shots from th

Strany 26 - 3.8.4 Tile Vertical

3-16 Software Framework PowerVista/312 User’s Manual3.11 Graphical Result Window FeaturesGraphical results presented in the graphical Result Wind

Strany 27 - 3.8.5 Arrange Icons

PowerVista/312 User’s Manual Software Framework 3-17• Autoscale: Via the Plot Context Menu, Autoscale toggles on and off the autoscaling feature

Strany 28 - 3.9.2 Phasing Toolbar

3-18 Software Framework PowerVista/312 User’s ManualNote: The Input Level Meter is displayed and hidden by successive clicks on its toolbar button

Strany 29 - 3.9.3 Status Bar

PowerVista/312 User’s Manual Device Configuration 4-14 Device Configuration4.1 Signal ConnectionsWARNING !ELECTRIC SHOCK HAZARD! Use proper el

Strany 30 - 3.10 Database View

4-2 Device Configuration PowerVista/312 User’s ManualPower system measurements always begin with a connection of external voltages to hardware inp

Strany 31 - 3.10.2 Column Descriptions

PowerVista/312 User’s Manual Device Configuration 4-34.1.2 Connection ProcedureIt is recommended that the following connection procedure always be

Strany 32

4-4 Device Configuration PowerVista/312 User’s Manual4.1.3 Three Phase Grounded WyeSystem Configuration PT Connection: WyeTo accurately measure a

Strany 33 - 3.12 Input Level Meter

PowerVista/312 User’s Manual Device Configuration 4-54.1.4 Three Phase UnGrounded WyeSystem Configuration PT Connection: Wye or DeltaThe three pha

Strany 34 - 3.13 Help

iiPowerVista/312 User’s ManualTable of Contents1 Introduction and Hardware Setup1.1 What is PowerVista/312 ……1-11.1.1 What Makes Up the Hardware ……1

Strany 35 - 4 Device Configuration

4-6 Device Configuration PowerVista/312 User’s Manual4.1.5 Three Phase DeltaSystem Configuration PT Connection: DeltaA three phase Delta connecti

Strany 36 - 4.1.1 Differential Inputs

PowerVista/312 User’s Manual Device Configuration 4-74.1.6 Three Phase Open Delta SystemSystem Configuration PT Connection: Open DeltaA three phas

Strany 37 - 4.1.2 Connection Procedure

4-8 Device Configuration PowerVista/312 User’s Manual4.1.8 120/240V ServiceSystem Configuration PT Connection: WyeA 120/240V Service connection a

Strany 38

PowerVista/312 User’s Manual Device Configuration 4-9Items in the System Configuration Dialog include:• Voltage Range (V1,2,3): is the voltage in

Strany 39

4-10 Device Configuration PowerVista/312 User’s Manual• V4 Voltage Range: is the voltage input range for channel V4. This setting allows the us

Strany 40 - 4.1.5 Three Phase Delta

PowerVista/312 User’s Manual Device Configuration 4-11Items in the Site Information Dialog include:• Site: is a 60 character text string that may

Strany 41 - 4.1.7 Single Phase

4-12 Device Configuration PowerVista/312 User’s Manual4.5 CommunicationCommunication settings define which hardware device EasyPower Measure is ac

Strany 42 - 4.2 System Configuration

PowerVista/312 User’s Manual Device Configuration 4-13Figure 4-13. Calibration Configuration Dialog.Items in the Calibration Configuration Dialog

Strany 43

4-14 Device Configuration PowerVista/312 User’s Manual Modifications to power calculations are performed as follows:⇒ Fundamental Phasor Power:

Strany 44 - 4.3 Site Information

PowerVista/312 User’s Manual Phaser Diagram 5-15 Phasor Diagram5.1 Phasor Diagram Overview The Phasor Diagram is a full featured measurement

Strany 45 - 4.4 Preferences

PowerVista/312 User’s Manualiii6 Detailed Harmonics6.1 Detailed Harmonics Overview ….. 6-16.2 Detailed Harmonics Configuration……6-16.3 Detailed

Strany 46 - 4.6 Calibration

5-2 Phaser Diagram PowerVista/312 User’s Manual• Harmonics View: determines which tabulation will be displayed in the Phasor Harmonics Window.•

Strany 47

PowerVista/312 User’s Manual Phaser Diagram 5-3Figure 5-2. 0.95 pf Lagging Wye Connection. Figure 5-3. 0.95 pf Lagging DeltaConnection.Figure

Strany 48

5-4 Phaser Diagram PowerVista/312 User’s Manual5.3.2 Total vs. Displacement Power FactorThe Phasor Demand Window displays all fundamental phasor

Strany 49 - 5 Phasor Diagram

PowerVista/312 User’s Manual Phaser Diagram 5-55.3.4 Demand TabulationThe Demand Tabulation Window is labeled to be self-descriptive. Several

Strany 50 - 5.3.1 Phasing

5-6 Phaser Diagram PowerVista/312 User’s ManualNotes:

Strany 51 - Connection

PowerVista/312 User’s Manual Detailed Harmonics 6-16 Detailed Harmonics6.1 Detailed Harmonics Overview Detailed Harmonics captures 10 cycles o

Strany 52 - 5.3.3 Harmonics Tabulation

6-2 Detailed Harmonics PowerVista/312 User’s ManualFigure 6-1. Detailed Harmonics Configuration Dialog.Items in the Detailed Harmonics Configura

Strany 53 - 5.3.4 Demand Tabulation

PowerVista/312 User’s Manual Detailed Harmonics 6-3⇒ Continuous Refresh: In this mode, as quick as a capture is processed and displayed, another

Strany 54

6-4 Detailed Harmonics PowerVista/312 User’s Manual6.3.3 Maximized SpectraThe maximizing feature of Detailed Harmonics is clearly useful when mea

Strany 55 - 6 Detailed Harmonics

PowerVista/312 User’s Manual Detailed Harmonics 6-56.3.5 Harmonics TabulationThe harmonics result tabulation in the Harmonics Table supplies both

Strany 56

ivPowerVista/312 User’s Manual

Strany 57 - 6.3.1 Average Harmonics

6-6 Detailed Harmonics PowerVista/312 User’s ManualNotes:

Strany 58 - 6.3.4 IEEE 519

PowerVista/312 User’s Manual Spectrum Analyzer 7-17 Spectrum Analyzer7.1 Spectrum Analyzer Overview The Spectrum Analyzer performs spectral deco

Strany 59 - 6.3.5 Harmonics Tabulation

7-2 Spectrum Analyzer PowerVista/312 User’s Manual• Input Channel: is the actual input channel (V1, V2, … I3, E1) to the hardware. This routine

Strany 60

PowerVista/312 User’s Manual Spectrum Analyzer 7-3V. We could include more frequencies left and right of the signal, but in real life those frequ

Strany 61 - 7 Spectrum Analyzer

7-4 Spectrum Analyzer PowerVista/312 User’s ManualChannel V1 SpectrumFrequency (Hz) Volts 73.242 1623.5 383.3 693.4 1003.4 1313.5 0.00

Strany 62 - 7.3.1 Hamming Window

PowerVista/312 User’s Manual Cycle-by-Cycle Capture 8-18 Cycle-by-Cycle Capture8.1 Cycle-by-Cycle Overview Cycle-by-Cycle Capture utilizes th

Strany 63

8-2 Cycle-by-Cycle Capture PowerVista/312 User’s Manual• Update Rate: is the rate at which the Cycle-by-Cycle graphic windows are updated. Note

Strany 64 - Channel V1 Spectrum

PowerVista/312 User’s Manual Cycle-by-Cycle Capture 8-3This method of display however can be confusing in terms of what data has actually been col

Strany 65 - 8 Cycle-by-Cycle Capture

8-4 Cycle-by-Cycle Capture PowerVista/312 User’s ManualFor arc furnace applications, poor SVC response can manifest itself through increased flick

Strany 66

PowerVista/312 User’s Manual Event Capture 9-19 Event Capture9.1 Event Capture Overview Event Capture continuously monitors all hardware channel

Strany 67

PowerVista/312 User’s Manual Introduction and Hardware Setup 1-11 Introduction and Hardware Setup1.1 What is PowerVista/312 ?PowerVista/312 is

Strany 68 - 8.3.4 Flicker

9-2 Event Capture PowerVista/312 User’s ManualFigure 9-1. Event Capture Configuration Dialog.Items in the Cycle-by-Cycle Configuration include:•

Strany 69 - 9 Event Capture

PowerVista/312 User’s Manual Event Capture 9-3• THD Trigger: is an event trigger that is compared against the voltage and current THD in percent

Strany 70

9-4 Event Capture PowerVista/312 User’s Manual• Monitored Harmonics: are the four user defined harmonics (from 1 to 50) that will be logged at t

Strany 71

PowerVista/312 User’s Manual Event Capture 9-5location of events at the bottom of the plot. When the cursor moves over a marker location, the cur

Strany 72

9-6 Event Capture PowerVista/312 User’s Manual9.3.3 Demand Log Data RecoveryIf the PC crashes from within Win95, or if there is a loss of power s

Strany 73

PowerVista/312 User’s Manual Frequency Following 10-110 Frequency Following10.1 Frequency Following OverviewFrequency Following is an EasyPowe

Strany 74

10-2 Frequency Following PowerVista/312 User’s Manual10.2.1 Frequency Following Process TimeFrequency Following is an additional algorithm, and

Strany 75 - 10 Frequency Following

PowerVista User’s Manual Glossary 11-1GlossaryCBEMA CurveA set of curves developed by the Computer Business Equipment Manufacturers Association (C

Strany 76

Glossary11-2 Glossary PowerVista/312 User’s ManualFourier TransformThe continuous Form of the Fourier Series is shown below in equation form.

Strany 77 - Glossary

GlossaryPowerVista User’s Manual Glossary 11-3K-FactorA component of the ANSI Standard C57.110-1986 transformer derating equation, and more common

Strany 78 - IT Product

1-2 Introduction and Hardware Setup PowerVista/312 User’s Manual1.1.2 What Is EasyPower Measure ?EasyPower Measure is a single software applic

Strany 79

Glossary11-4 Glossary PowerVista/312 User’s ManualResonanceA condition where elements of a circuit match impedance to form a total high impedance

Strany 80

GlossaryPowerVista User’s Manual Glossary 11-5Total Harmonic Distortion (THD)The total "root sum square" harmonic distortion in percent

Strany 81 - Transformer Derating Factor

Glossary11-6 Glossary PowerVista/312 User’s ManualWatts and VarsFundamental Watts and Vars are defined from pre-calculated voltage and current vec

Strany 82 - Watts and Vars

PowerVista/312 User’s Manual Introduction and Hardware Setup 1-3Figure 1-2. PowerVista/312, Hardware Identification