21
Chapter 4
Specifications
All specifications are subject to change without notice.
Typical for 25 °C unless otherwise specified.
All specifications apply to all temperature and voltage input channels unless otherwise specified.
Specifications in italic text are guaranteed by design.
Analog input
Table 1. General analog input specifications
ADS1256, 24-bit Sigma Delta
3750 S/s, 2000 S/s, 1000 S/s, 500 S/s, 100 S/s, 60 S/s, 50 S/s,
25 S/s, 10 S/s, 5 S/s, 2.5 S/s
Single channel: 2.5 S/s to 1102.94 S/s (software-selectable)
Multiple channels: 0.16 S/s to 551.47 S/s (software-selectable)
See Table 11 and Table 12 for details.
16 single-ended or 8 differential (software-selectable).
Thermocouples require differential mode. For each channel
configured as differential, you essentially lose two single-ended
channels.
500 VDC min between field wiring and USB interface
Temperature sensor input, software-selectable to match sensor
type
±10 V, ±5 V, ±2.5 V, ±1.25 V, ±0.625 V, ±0.3125 V,
±0.15625 V, ±0.078125 V (software-selectable)
Absolute maximum input
voltage
±22 V max (power on)
±10 V max (power off)
10 MΩ (power on)
390 Ω (power off)
Input voltage >±30 V
(power on/off)
Maximum working
voltage (signal + common
mode)
Common mode rejection
ratio (Note 1)
Thermocouple mode,
(f
IN
= 60 Hz)
Voltage mode,
(f
IN
= 60 Hz, all input
ranges)
Software-selectable channel and range
Software-selectable for each channel.
Komentáře k této Příručce