O - Recall Configuration
SYNTAX
Oval Load configuration val, where val = value from 0 to 100.
O? Returns currently loaded configuration val.
DESCRIPTION
The Recall Configuration (
O) command is used to load a previously-saved configuration from Non-Volatile RAM and
use it as the present configuration. The argument of the command
val may range from 0 to 100. Configuration 0 is
the power-up configuration. Any load of a configuration not previously-saved with the Save Configuration (
S)
command, configures the channel with the default setting for each command (i.e. Data Format
F0, Bus Terminator Y0,
etc.) Issuing an O? query command returns the configuration number last recalled.
EXAMPLE
PRINT#1,"CLEAR08"
Line 1: Reset the Digital488/80A..
PRINT#1,"OUTPUT08;O5X"
Line 2: Load Configuration 5 from Non-Volatile RAM for use as
the present configuration.
Digital488/80A User’s Manual
967695 Digital488/80A Commands 63