Section 3 Command Descriptions
3.14
Invert In
The
Invert
command is used to change the polarity of the handshake and data
lines. At power up all handshake and control lines are active high (logic one = + 5
volts). The
Invert
command can selectively change the polarity of each of the
handshake lines, and of the data lines. If multiple
Invert
commands are contained
within the same string, then an
Execute
command (X) should be included between
each
Invert
command. An alternative is to add the values of each
Invert
command
desired, and send one command with the sum of the desired commands. The
Invert
commands are Ored together as received. To delete any one command, it is necessary
to program the default mode
I0
, then re-program the desired commands.
I0
All control lines are active high, all data lines are high true.
I1
Inhibit
output
is active low
I2
Trigger
output
is active low
I4
Data Strobe
output
is active low
I8
Clear
output
is active low
I16
Data
is low
true
I32
EDR
input
is falling-edge sensitive
I64
Service
input
is falling-edge sensitive
Example:
@
reset the Digital 232
I32XI64
select EDR and Service
input
as
falling-edge sensitive
note:
I96 performs the same function as above