Section 3 Command Descriptions
3.20
Invert In
The
Invert
command is used to change the polarity of the handshake and data
lines. At power up all handshake and control lines are active high (logic one = + 5
volts). The
Invert
command can selectively change the polarity of each of the
handshake lines, and of the data lines. If multiple
Invert
commands are contained
within the same string, then an
Execute
command (
X
) should be included between
each
Invert
command. An alternative is to add the values of each
Invert
command
desired, and send one command with the sum of the desired commands. The
Invert
commands are ORed together as received. To delete any one command, it is necessary
to program the default mode
I0
, then reprogram the desired commands.
I0
All control lines are active high, all data lines are high true
I1 Inhibit
output is active low
I2 Trigger
output is active low
I4 Data Strobe
output is active low
I8 Clear
output is active low
I16 Data
is low true
I32 EDR
input is falling-edge sensitive
I64 Service
input is falling-edge sensitive
Example:
CLEAR 718
reset the
Digital488
OUTPUT 718;"I32XI64X"
select EDR and Service input as
falling-edge sensitive
Note:
OUTPUT 718;"I96X"
performs the same function as above