Section 4 Serial Controller
4.19
ARM
The
ARM
command programs the Analyzer488 to send event messages to the
serial host when one or more specified events (
SRQ
,
TRIGGER
,
RECORD
, or
STEP
)
occur. The event messages returned are the same strings used to program the events.
Once a condition is
ARMed
it remains
ARMed
until the event specified has
occurred, it is
DISARMed
, or the Analyzer488 is reset. The
ARM
command must be
resent after an event message is reported to reactivate the
ARM
ed condition.
There are two types of events, level sensitive and edge sensitive. The level
sensitive event,
SRQ
, will be reported immediately if the SRQ line is asserted when
ARM
is issued. Usually, some action must be taken, such as issuing a
SPOLL
command to clear SRQ, prior to reissuing the
ARM
command. The edge sensitive
events,
TRIGGER, RECORD, and STEP
, are cleared when reported.
The query option,
ARM?
, returns the conditions that are presently
ARM
ed.
TYPE System Command
SYNTAX ARM
[[?]|[event[event…]]]
event
is one or more of
SRQ
,
TRIGGER,
RECORD,
or
STEP
If no event is specified,
ARM
SRQ
is assumed.
RESPONSE Event string sent when the event occurs
REMARKS The following conditions are available using the ARM command:
SRQ
The event message '
SRQ
' is returned to the serial host when the state
of the Service Request (SRQ) bus line is detected to be in the
asserted state. This event is level sensitive. If the condition exists at
the time the
ARM
SRQ
command is issued, the Analyzer488 will
return the event message immediately. If the
ARM
command is issued
without any specified events,
ARM
SRQ
is assumed.