DAQFlex Software User's Guide DAQFlex Message Reference Programming messages AI
33
Get the status of the A/D internal calibration.
Message
"?AI:ADCAL/STATUS"
Response
"AI:ADCAL/STATUS=value"
value
RUNNING, IDLE
CAL
Enable or disable calibration of all A/D channels.
Message
"AI:CAL=value"
Response
"AI:CAL"
value
ENABLE, DISABLE
Example
"AI:CAL=ENABLE"
Note
This message is processed by the DAQFlex Software library, and is not sent to the
device.
Get a value indicating whether the calibration coefficients will be applied to the raw A/D data.
Message
"?AI:CAL"
Response
"AI:CAL=value"
value
ENABLE, DISABLE
Note
This message is processed by the DAQFlex Software library, and is not sent to the
device.
CJC
Get the CJC value in the specified format.
Message
"?AI{ch}:CJC/format"
Response
"AI{ch}:CJC/format=value"
format
DEGC, DEGF, KELVIN
value
The measured temperature.
Example
"?AI{0}:CJC/DEGC"
CHMODE
Set the analog input mode to single-ended or differential for all channels.
Message
"AI:CHMODE=value"
Response
"AI:CHMODE"
value
SE, DIFF
Example
"AI:CHMODE=SE"