WEB-TEMP User's Guide Specifications
33
Note 11: Error shown does not include errors of the sensor itself. The sensor linearization is performed
using a Steinhart-Hart linearization algorithm. The accuracy and tempco specifications include the
accuracy of the Callendar-Van Dusen linearization algorithm. These specifications are for one year
while operation of the device is between 15 °C and 35 °C. The specification does not include lead
resistance errors for 2-wire thermistor connections. Contact your sensor supplier for details on the
actual sensor error limitations. Total thermistor resistance on any given channel pair must not exceed
180 kΩ. Typical resistance values at various temperatures for supported thermistors are shown in Table
8.
Table 8. Typical thermistor resistance specifications
Note 12: Resistance values greater than 180 kΩ cannot be measured by the WEB-TEMP in the thermistor
mode. The 180 kΩ resistance limit includes the total resistance across the current excitation (±Ix) pins,
which is the sum of the thermistor resistance and the lead resistances.
Note 13: For accurate three wire compensation, the individual lead resistances connected to the ±Ix pins
must be of equal ohmic value. To ensure this, use connection leads of equal lengths.
Throughput rate
Table 9. Throughput rate specifications
2 S/s on each channel, 4 S/s total
2 S/s on each channel, 6 S/s total
2 S/s on each channel, 8 S/s total
2 S/s on each channel, 10 S/s total
2 S/s on each channel, 12 S/s total
2 S/s on each channel, 14 S/s total
2 S/s on each channel, 16 S/s total
Note 14: The analog inputs are configured to run continuously. Each channel is sampled twice per second.
The maximum latency between when a sample is acquired and the temperature data is provided by the
Ethernet unit is approximately 0.5 seconds.