USB-TC-AI User's Guide Specifications
19
Analog input calibration
Table 9. Analog input calibration specifications
+10.000 V, ±5 mV max. Actual measured values stored in EEPROM
Long term stability: 30 ppm/1000 h
Throughput rate
Table 10. Throughput rate specifications
2 S/s on each channel, 4 S/s total
2 S/s on each channel, 6 S/s total
2 S/s on each channel, 8 S/s total
2 S/s on each channel, 10 S/s total
2 S/s on each channel, 12 S/s total
2 S/s on each channel, 14 S/s total
2 S/s on each channel, 16 S/s total
Note 11: The analog inputs are configured to run continuously. Each channel is sampled twice per second.
The maximum latency between when a sample is acquired and the voltage/temperature data is
provided by the USB unit is approximately 0.4 seconds.
Digital I/O
Table 11. Digital input/output specifications
Independently configured for input or output.
Power on reset is input mode.
Pull-up/pull-down
configuration
All pins pulled up to +5 V via 47 K resistors (default). Contact MCC factory for pull-
down to ground (GND) capability.
Digital I/O transfer rate
(software paced)
Digital input: 50 port reads or single bit reads per second typ
Digital output: 100 port writes or single bit writes per second typ
2.0 V min, 5.5 V absolute max.
0.8 V max, –0.5 V absolute min
Output low voltage
(IOL = 2.5 mA max)
Output high voltage
(IOH = -2.5 mA max)
Note 12: All ground pins on the USB-TC-AI (pins 9, 19, 22, 27, 30, 33, 36, 39, 49) are common and are
isolated from earth ground. If a connection is made to earth ground when using digital I/O and
conductive thermocouples, the thermocouples are no longer isolated. In this case, thermocouples must