Chapter 4
Specifications
All specifications are subject to change without notice.
Typical for 25°C unless otherwise specified.
Specifications in italic text are guaranteed by design.
Analog input
Table 1. General analog input specifications
Parameter
Condition
Specification
A/D converter type
16-bit successive approximation type
Number of channels
8 single-ended
Input configuration
Individual A/D per channel
Sampling method
Simultaneous
Absolute maximum input voltage
CHx IN relative to GND
±15 V max
Input impedance
100 MΩ min
Input ranges
Software-selectable per channel
±10 V, ±5 V, ±2 V, ±1 V
Sampling rate
Hardware paced
0.01 S/s to 100 kS/s, software-selectable
Throughput
Software paced
500 S/s all channels
Hardware paced (Note 1)
(400 kS/s) / (# of channels) max,
100 kS/s max for any channel
BURSTIO ≤ 32,768 total samples
(uses onboard FIFO)
(800 kS/s) / (# of channels) max,
100 kS/s max for any channel
Gain queue
Software configurable. Up to eight
elements one gain element per unique,
ordered channel.
Resolution
16 bits
No missing codes
16 bits
Crosstalk
Signal DC to 25 kHz
80 dB
Trigger source
Software-selectable
External digital: TRIG_IN
Note 1: Maximum throughput when scanning in hardware paced mode is machine dependent.
Accuracy
Analog input DC voltage measurement accuracy
Table 2. Calibrated absolute accuracy
Range
Accuracy (mV)
±10 V
5.66
±5 V
2.98
±2 V
1.31
±1 V
0.68