When the
Serial488A
is addressed to listen, it accepts data from the
active talker and outputs this data through the serial interface. It substitutes
the selected serial terminators for the received IEEE bus terminators.
4.6.3 Device Clear
(DCL and SDC)
Device Clear resets the
Serial488A
's IEEE input and serial input
buffers. Any pending data and Service Requests (
SRQ
), including the
information they convey, are lost.
4.6.4 Interface Clear
(IFC)
IFC places the
Serial488A
in the Talker/Listener Idle State. It clears
any pending requests for service (SRQ). The condition which caused the
SRQ remains unmodified.
4.6.5 Serial Poll Enable
(SPE)
When Serial Poll Enabled, the
Serial488A
sets itself to respond to a
serial poll with its serial poll status byte if addressed to talk. When the
serial poll byte is accepted by the controller, any pending
SRQ
s are
cleared. The
Serial488A
will continue to try to output its serial poll
response until it is 'Serial Poll Disabled' by the controller.
4.6.6 Serial Poll Disable
(SPD)
Disables the
Serial488A
from responding to serial polls by the
controller.