Measurement-computing DAC488 v.1 Uživatelský manuál Strana 27

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 114
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 26
DAC488 User’s Manual DAC488 Power & Assembly 21
To Verify the Port Voltage – 10-Volt Range
1. Send the following command string to the DAC488:
C0 P1 A0 R3 V10 X where:
C0
= Direct Control Mode
R3
= 10-Volt Range
P1 = Port 1 V10 = Output +10 Volts
A0 = Autorange Off X = Execute
2. Send the following command string to the 199 DMM:
F0 R3 Z0 N1 S1 X where:
F0 = DC Volts N1 = Scanner Channel 1
R3
= 30-Volt Range
S1
= 5 ½ Digit Resolution
Z0 = Zero Disabled X = Execute
3. For each test, the specific Value Output (
V) command, its commanded output voltage, and its
corresponding Low Limit and High Limit, are as follows:
V10 = Output +10 V +9.985 Reading +10.015 Volts
V5 = Output +5 V +4.990 Reading +5.010 Volts
V1 = Output +1 V +0.994 Reading +1.006 Volts
V0 = Output 0 V -0.005 Reading +0.005 Volts
V-1 = Output -1 V -1.006 Reading -0.994 Volts
V-5 = Output -5 V -5.010 Reading -4.990 Volts
V-10 = Output -10 V -10.015 Reading -9.985 Volts
To Verify the Port Voltage – 5-Volt Range
1. Send the following command string to the DAC488:
C0 P1 A0 R2 V5 X where:
C0 = Direct Control Mode R2 = 5-Volt Range
P1 = Port 1 V5 = Output +5 Volts
A0 = Autorange Off X = Execute
2.
Send the following command string to the 199 DMM:
F0 R3 Z0 N1 S1 X
where:
F0 = DC Volts N1 = Scanner Channel 1
R3
= 30-Volt Range
S1
= 5 ½ Digit Resolution
Z0 = Zero Disabled X = Execute
3.
For each test, the specific Value Output (
V
) command, its commanded output voltage, and its
corresponding Low Limit and High Limit, are as follows:
V5 = Output +5 V +4.9920 Reading +5.0080 Volts
V2.5 = Output +2.5 V +2.4945 Reading +2.5055 Volts
V1
= Output +1 V +0.9960 Reading +1.0040 Volts
V0 = Output 0 V -0.0030 Reading +0.0030 Volts
V-1 = Output -1 V -0.9960 Reading -1.0040 Volts
V-2.5 = Output –2.5 V -2.4945 Reading -2.5055 Volts
V-5 = Output -5 V -4.9920 Reading -5.0080 Volts
Zobrazit stránku 26