3.3.5 Common Mode Voltage > +/-10V
The CIO-DAS16/F will not directly monitor signals with common mode voltages greater than +/-10V. You will either
need to alter the system ground configuration to reduce the overall common mode voltage, or add isolated signal
conditioning between the source and your board (Figure 3-8).
Figure 3-8. Common Mode Voltage > +/-10V - Differential Input
3.3.6 Isolated Grounds / Single-Ended Inputs
Single-ended inputs can be used to monitor isolated inputs, though the use of the differential mode will increase you
system’s noise immunity. Figure 3-9 shows the recommended connections is this configuration.
Figure 3-9. Isolated Grounds / Single-Ended Inputs
16
System with a Large Common Mode Voltage,
Connected to a Differential Input
L
arge common
mode voltage
between signal
source & A/D board
GND
Isolation
Barrier
When the volta
g
e difference
between si
g
nal source and
A/D board
g
round is lar
g
e
enou
g
h so the A/D board’s
common mode ran
g
e is
exceeded, isolated si
g
nal
conditionin
g
must be added.
+
-
Input
Amp
To A/D
A/D Board
I/O
Connector
LL GND
CH Hi
g
h
CH Low
10 K
10K is a recommended value. You may short LL GND to CH Low
instead, but this will reduce your system’s noise immunity.
Isolated Si
g
nal Source
Connected to a Sin
g
le-Ended Input
I/O
Connector
+
-
Input
Amp
To A/D
LL GND
CH IN
A/D Board
Isolated
signal
source