Measurement-computing CB-COM-7013 Uživatelský manuál Strana 22

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 44
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 21
22
CB-7013, CB-7033 Users Manual
2.5 $AA0
Description: Perform Span Calibration
Syntax: $AA0[CHK](cr)
$ Delimiter character
AA Address of setting module (00 to FF)
0 Command for span calibration
Response: Valid Command: !AA[CHK](cr)
Invalid Command: ?AA[CHK](cr)
Syntax error or communication error may get no re-
sponse.
! Delimiter for valid command.
? Delimiter for invalid command.
AA Address of response module (00 to FF).
Example:
Command: $010 Receive: !01
Perform address 01 span calibration, return successful.
Command: $020 Receive: ?02
When performing address 02 zero calibration, return was not
enabled before performing calibration command.
Related Command:
Sec2.6 $AA1, Sec. 2.15 ~AAEV
Related Topics:
Sec. 1.8 Calibration
Zobrazit stránku 21